Catheter TypeRapid Exchange
Proximal Shaft1.9F (Ø 1.5 mm); 2.1F (Ø1.75 mm – 4.0 mm)
Distal Shaft2.36F – 2.7F (Ø1.5 mm – 4.0 mm)
Catheter Working Length140 cm
Tip Length11.5 mm – 2.5 mm (Ø1.5 mm – 4.0 mm)
Marker (s)1 (Ø1.5 mm); 2 (Ø1.75 mm – 4.0 mm)
Crossing Profile0.0313” (Ø3.0 mm)
CoatingHydrophilic (distal tip to guidewire exit port); Hydrophobic (guidewire lumen)
Nominal Pressure (NOM) 16 atm (Ø1.5 mm); 14 atm (Ø1.75 mm – 4.0 mm)
Rated Burst Pressure (RBP)16 atm
Balloon MaterialNylon (Ø1.5 mm); Pebax (Ø1.75 mm – 4.0 mm)
Balloon Folds3
Balloon Diameters (mm)1.5, 1.75, 2.0, 2.25, 2.5, 2.75, 3.0, 3.25, 3.5, 4.0
Balloon Lengths (mm)10, 12, 15, 20, 30
Guiding Catheter Compatibility5F
Key Design Features (for all sizes) Sapphire 3 Schematic

Compliance Chart

Pressure (atm) Balloon Diameter (Ø mm)
1.5 1.75 2.0 2.25 2.5 2.75 3.0 3.25 3.5 4.0
2 1.42 1.66 1.87 2.11 2.36 2.59 2.82 3.06 3.29 3.77
4 1.46 1.70 1.94 2.18 2.43 2.67 2.91 3.16 3.39 3.89
6 NOM* 1.50 1.75 2.00 2.25 2.50 2.75 3.00 3.25 3.50 4.00
8 1.54 1.80 2.06 2.32 2.57 2.83 3.09 3.34 3.61 4.11
10 1.58 1.84 2.13 2.39 2.64 2.91 3.18 3.44 3.71 4.23
12 1.62 1.89 2.19 2.45 2.71 2.98 3.27 3.53 3.82 4.34
14 RBP** 1.66 1.94 2.25 2.52 2.78 3.06 3.36 3.62 3.92 4.45
16 RBP** 1.70 1.98 2.32 2.59 2.85 3.14 3.45 3.72 4.03 4.57
18 1.74 2.03 2.38 2.66 2.92 3.22 3.54 3.81 4.14 4.68
20 1.78 - - - - - - - - -